G2500 Medical Gas Outlet Analyzer Specifications

Tests Performed

Types of Outlets Tested

Static pressure

Medical Air

    Maximum static pressure:  220 psig

Nitrogen

Dynamic pressure at specified flow rates

Nitrous Oxide

    Maximum flow rate:  160 LPM

Oxygen

Pressure drop

Vacuum

Gas Type Determination

 

Data Collection Capability

Stores approximately 1,500 outlet test results.

Reports

Generate reports on an IBM® compatible computer and your printer.

Reports show every outlet tested with area, room and outlet number identified, along with the following information:

  • Type of gas expected and identified (expected gas type is determined by the specific gas adapter attached to the analyzer)
  • Static pressure of outlet
  • Restricted flow rate of outlet and the pressure at that flow rate
  • Overall evaluation and status of outlet as OK or BAD
  • User Comments
  • Notes entered by operator at individual outlets

Options for report formats include listing of all outlets tested or only those that need attention.

Merge individual files into one report.

Detected Gas Type

The detected gas type is determined by measuring the percent of Oxygen and percent of Nitrous Oxide in the gas sample being tested.  The table below illustrates the methodology used to determine a specific gas type.

GAS TYPE

PRESSURE

% OXYGEN

% NITROUS OXIDE

Oxygen

> 4 PSIG

70 - 100

0 - 30

Nitrous Oxide

> 4 PSIG

0 - 10

70 - 100

Medical Air

> 4 PSIG

10 - 30

0 - 30

Nitrogen

> 4 PSIG

0 - 10

0 - 30

Vacuum

> 1 inHg

NEG P

NEG P

Unknown

Any gas not falling within the above parameters

 

 